Environmental | |
Operating temperature | -10 °C to +50 °C |
Storage temperature | -20 °C to +60 °C |
Power requirements | 12 V dc (Universal ac adapter/charger supplied) |
Battery | Internal 3800 mAh advanced NiMH pack |
Operating time | |
On full charge | Approx 50 hours (Measure only or external air; no pump) |
Using internal pump | Approx 100 calibration cycles to 300 psi |
Physical | |
Dimensions | 15.25 in L x 12 in W x 7 in D |
Weight | ~7 kg (15 lb) |
EMI/RFI conformance | EN61326:2006 Annex A |
Connectors/ports | 1/8 in NPT (External supply port and test port) |
Included accessories | Manual, NIST-traceable certificate, test leads, universal ac adapter/charger |
Ranges | |
Pressure (internal pump) | -80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar) |
Pressure (external air) | 0 to 2 MPa (0 to 300 psi, 0 to 20 bar) |
mA | 0 to 24.000 mA |
Volts | 0 to 30.000 V dc |
Engineering units | psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4 °C, mmH2O @ 20 °C, cmH2O @ 4 °C, cmH2O @ 20 °C, inH2O @ 4 °C, inH2O @ 20 °C, inH2O @ 60 °F, mmHg @ 0 °C, inHg @ 0 °C |
Instrumental measurement uncertainty | |
Pressure | ±(0.025 % of reading + 0.01 % FS) |
mA | ±(0.015 % of reading + 0.002 mA) |
Volts | ±(0.015 % of reading + 0.002 V) |
Temperature effect (all functions) | |
No effect on accuracy on all functions from 15 °C to 35 °C | |
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C |